Lambda Transmitter LT2

LT2
  • Universal O2 measuring
  • Semi- and fully automatic calibration
  • Applicable up to 450 °C
  • Detection of CO/H(COequivalent)
  • Ocontrol integrated

Measuring accuracy
(with Lambda Probe LS2)

+/- 10% of measured value 
max. +/-0,3% by vol. O2

Settling time (90% time)

T90 < 15s

Ambient temperature
– Operation 
– Transport and storage

-20°C…+60°C
-40°C…+85°C

Power supply

230 V AC und 115 V AC
+10% / -15%, 48 Hz…62Hz

Power consumption

Typically 50 VA, short-term 150 VA (probe heating-up phase)

Time for operational status 
to be achieved with

approx. 10 minutes after “POWER ON”